Waltron 9065 Quick Reference Manual page 10

Luminescent dissolved oxygen analyzer calibration kit
Hide thumbs Also See for 9065:

Advertisement

Step
3.6
3.7
3.7
104-001-A.7 Released 06/14/2019
Description
After approximately 2 minutes the
analyzer will show an alert for
successful calilbration.
Press any button to clear this
alert.
Close the valve on the gas
regulator to preserve your
calibration gas.
Use the arrow buttons to select
STORE CALIBRATION. Press
ENTER button.
Wait for approximately 2 minutes
the analyzer will show an alert for
successful storage of calibration.
Key Points
If calibration is not successful,
make sure the sensor spot is
completely dry and try again.
Contact Waltron technical
support if you are still
unsuccessful.
Perform this step to store the
calibration results. Otherwise
the analyzer will revert back to
the last calibration results and
the calibration process must be
repeated.