Waltron 9065 Quick Reference Manual page 5

Luminescent dissolved oxygen analyzer calibration kit
Hide thumbs Also See for 9065:

Advertisement

Step
2
CONNECT THE NITROGEN CALIBRATION GAS
2.1
2.2
2.3
104-001-A.7 Released 06/14/2019
Description
Attach M5 barb fitting (provided)
to calibration beaker inlet.
The inlet is the side that is closer
to the flange than the other
fitting.
A calibration beaker is included
with the initial analyzer order.
Install the calibration beaker by
pushing straight upwards over the
sensor.
Inspect O-Ring on provided gas
regulator for any debris or
damage.
Key Points
Ensure the provided O-Ring is
installed on the fitting to
provide a seal against leaks
between the fitting and the
beaker.
Take care to ensure there
are no leaks between the
calibration gas tank and the
sensor.