Luminescent dissolved oxygen analyzer calibration kit
Hide thumbs Also See for 9065:

Advertisement

Step
1.5
1.6
1.7
1.8
104-001-A.7 Released 06/14/2019
Air dry
Description
Disconnect the plumbing from the
flow cell.
Remove the flow cell.
Allow the sensor to air dry and to
allow the sensor and the
calibration kit to equalize to the
same temperature. You may flow
extra calibration gas or clean dry
air over the sensor to expedite
drying.
Store the calibration kit in the
same room as the sensor for at
least 4 hours while the sensor is
drying to equalize to the same
room temperature.
Key Points
DO NOT wipe dry the black
sensor spot. Touching the
sensor may damage the
delicate coating.
The calibration process is
temperature sensitive. The
temperature of the sensor
must be roughly equal to the
temperature of the calibration
gas.