Fm / Us Hazloc Markings; Csa / Canada Hazloc Markings - ABB MT Series Manual

Guided wave radar level transmitter
Hide thumbs Also See for MT Series:
Table of Contents

Advertisement

3.3 FM / US HazLoc ratings
Factory Mutual Research Corporation
MT50xx, MT51xx and MT52xx f/b abcdefghiFMjk :
XP-IS / I / 1 / ABCD / T6 Ta = 77°C;
DIP / II,III / 1 / EFG / T6 Ta = 77°C; Type 4X
IS / I / 1 / CD / T4 Ta = 77°C – ELE9020; Entity;
Type 4X
NI / I / 2 / ABCD / T4 Ta = 77°C;
S / II,III / 2 / FG / T4 Ta = 77°C; Type 4X
ANI / I / 2 / ABCD / T4 – ELE9020; Entity; Type 4X
MT51xx and MT52xx f/b abcdefghiFMjk :
XP-IS / I / 1 / ABCD / T6 Ta = 77ºC
DIP / II, III / 1 / EFG / T6 Ta = 77ºC
IS / I / 1 / ABCD / T4 Ta = 77ºC - ELE1034
NI / I / 2 / ABCD / T4 Ta = 77ºC
S / II,III / 2 / FG / T4 Ta = 77ºC
ANI / I / 2 / ABCD / T4 - ELE1034 Type 4X
Conditions of Certification
(a) Special Conditions for safe use
For guidance on Intrinsically Safe U.S. installations, see AN-
SI/ISA-RP12.06.01, Installation of Intrinsically Safe Systems
for Hazardous (Classified) Locations.
Installations shall comply with the manufacturer's instruc-
tions, according to drawing ELE1034 or ELE9020 as applica-
ble.
Tampering and replacement with non-factory components
may adversely affect the safe use of the system.
Installations in the US shall comply with the relevant require-
ments of the National Electrical Code® (ANSI/NFPA-70
(NEC®).
(b). Conditions of Manufacture
The manufacturer shall supply copies of Control Drawing
ELE1034 & ELE9020 with each transmitter. The manufactur-
er shall make additional copies available upon request.
The manufacturer shall make any Special Conditions of Use
available to the user of the Product.
The manufacturer shall carry out verifications or tests nec-
essary to ensure that the electrical equipment produced
complies with the documentation.
3.4 CSA / Canadian HazLoc ratings
Canadian Standards Association
Model MT5000/5100/5200 Radar Level Transmitter
"[Ex ia] Associated Equipment"
« Explosion-proof with intrinsically safe antenna rod »
XP CL I, DIV 1, GP ABCD;
CL II, DIV 1, GP EFG; CL 3 - T6
CL I, DIV 2, GP ABCD;
CL II, DIV 2, GP EFG - T5
Maximum ambient = 77ºC.
MWP = 5000 psi max.
Enclosure Type 4X. Single Seal.
"[Ex ia] INTRINSICALLY SAFE"
« Intrinsically Safe Entity, when installed per ELE1034 »
IS CL I, DIV 1, GP CD;
CL II, DIV 1, GP EFG - T4
Maximum ambient = 77ºC.
MWP = 5000 psi max.
Enclosure Type 4X. Single Seal.
Conditions of Certification
(a) Special Conditions for safe use
For explosion-proof and Division 2 models with the symbol
"[Ex ia]" and the words "Associated Equipment", always
open circuit before removing covers.
For Intrinsically Safe models with the symbol "[Ex ia]" and
the words "INTRINSICALLY SAFE", installations shall comply
with the installation instructions, according to drawing
ELE1034.
Caution: substitution of components may impair suitability
for use as I.S. or Div. 2.
Caution: do not disconnect equipment unless power has
been switched off or the area is known to be non-hazardous
Installations in Canada shall comply with the relevant re-
quirements of the Canadian Electrical Code® (CEC).
(b). Conditions of Manufacture
The manufacturer shall supply copies of Control Drawing
ELE1034 with each transmitter. The manufacturer shall
make additional copies available upon request.
The manufacturer shall make any Special Conditions of Use
available to the user of the Product.
The manufacturer shall carry out verifications or tests nec-
essary to ensure that the electrical equipment produced
complies with the documentation.
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mt5000Mt5100Mt5200

Table of Contents