Intrinsic Safe Control Document; Intrinsically Safe Fm Us And Canada Approved - ABB DHH805-A Operating Instruction

Handheld configuration devices with hart communications
Table of Contents

Advertisement

9.2 Intrinsic Safe Control Document

9.2.1 Intrinsically safe FM Us and Canada Approved*

Hazardous classified location
- +
HART pressure transmitter
Entity Parameters
V max = 30 Vdc
I max = 215 mA
P max = 0.85 W
Ci = 0.04 uF
Li = 0 mH
Voc = 2.6 Vdc
Isc = 18 mA
Po = 61mW
Ca = 99 uF
La = 100mH
Intrinsically safe
FM us : CL 1 Div 1 Gr A B C D, T4
CL 1, Zone 0, IIC, T4
FM c : CL 1, Zone 0, IIC, T4
Warning. Use of the charger port is not allowed while the DHH805-A
is connected or located in hazardous area
*Note. For FMc the equipment is approved only for the Canadian Zone Concept.
46 OI/DHH805-EN Rev. D | Handheld communicator Model DHH805-A
Unclassified location
Barrier
Note: For FMc Approval, use the FMc barriers
DHH805A
Warning. Use of the USB port is not allowed while the DHH805-A is
connected or located in hazardous area
9 Hazardous area considerations
I.S.
load
250 ohm min.
Po Barrier < 0.7 Watts
Note: see paragraph 9.2.3 for Entity Concept
according to Canadian Zone Concept
30 V max , 330 ohms min.
28 V max , 300 ohms min.
22 V max , 180 ohms min.
Power
Supply

Hide quick links: