Using the Bed Scale

The Bed Scale is provided with the integrated hold function to
determine the average weight. It enables people to be weighed
accurately although they are not still on the bed.
Note: Determining average weight is not possible incase of big fluctuation
between weight.
Switch 'ON' the scale using [ON/OFF] key. The diagnostic self-
check will be performed and after that „0.0 kg" will display on
the screen with 'stable' and 'zero sign'.
Call the bed weight from Memory
and press ENTER key, the scale will Pre-tare the bed
section)
weight and (-) sign will appear in front of bed weight.
Move the bed (Patient is in bed) on the scale platform.
Press the [HOLD] key. The 'HOLD' will display on the screen
with blinking triangle and after few seconds the average of
fluctuating weight will display on the screen and will lock on the
display.
To release the locked weight on display simply press the
[HOLD] key again or remove the object/ person from the scale
platform and make sure there is no weight on platform: the
display will return to „0.0 kg" display automatically.
HOLD key function can be activated before or after putting the
weight on the tread platform. But in case of weighing unstable
person it is recommended to press HOLD key after the person
moves on the tread platform.
(Refer- Re-calling Pre-tare weight