Normal Weighing Procedure (Patient On The Bed) - Faaborg Rehab Technic Charder MS 6000 User Manual

Bed scale
Table of Contents

Advertisement

Normal Weighing Procedure (patient on the bed)

Turn On the scale using ON/OFF
key.
**(before moving bed on the
weigh bridge)
Call the PRE-TARE value (bed
weight)
from
memory
(refer
Recall PRE-TARE value section)
and then press Enter, scale will
automatically
perform
Tare
function and the minus (-) sign
will appear in front of Pre-Tare
value.
Place both the weigh bridges
under the bed and push the bed
on the scale.
The upper display will show the
net weight of the patient.
Now if you want BMI also, simply
long press HOLD/BMI key for 3
seconds, the middle display will
show the last height of the patient
and extreme digit will blink. Key in
the
height
measurement
of
patient using 0~9 key and press
ENTER key. The third display will
show the BMI of the patient
- 21 -

Advertisement

Table of Contents
loading

Table of Contents