Guidance and manufacturer's declaration-electromagnetic immunity
The MEDICAL SCALE MS6000 is intended for use in the electromagnetic
environment specified below.
The customer or the user of the MEDICAL SCALE MS6000 should assure that is
used in such and environment.
IEC 60601 test
Immunity test
level
Conducted RF
3 Vrms
IEC 61000-4-6
150 KHz to 80 MHz
3 A/m
3 A/m
Compliance
Electromagnetic
level
environment-guidance
3 Vrms
Portable and mobile RF
communications equipment
should be used no closer to any
part of the
MS6000 including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in metres (m).
10
- -
The MEDICAL SCALE
MS6000 power
frequency magnetic
fields should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.