CAUTION
• Clean and lubricate the handpiece prior to sterilizing. If blood remains on the internal surface it can become clotted and
cause product failure.
• Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
• To avoid product failure, do not use a sterilizer that exceeds a cycle temperature of 136°C, including the dry cycle. In some
sterilizers, the chamber temperature may exceed 136°C. Contact the sterilizer manufacturer for detailed information about
cycle temperatures.
• Do not touch the product immediately after steam sterilization as it will be very hot and must remain in a sterile condition.
• Steam sterilization is recommended for the product. The validity of other sterilization methods (such as plasma sterilization
or EOG sterilization) is not confirmed.
NOTICE
• NSK recommends Class B sterilizers as stated in EN 13060.