NOTICE
• NSK recommends the use of "Spray Mist Absorber" (Y900084) to prevent oil mist expelling out of the handpiece head.
Automatic Method (Cleaning and Lubricating the Interior)
Use an NSK automatic lubrication system to perform lubrication.
For information on using an automatic lubrication system, see the instruction manual for the automatic lubrication system.
WARNING
• If blood infiltrates inside a handpiece, an automatic handpiece cleaning and lubrication system
may not totally clean the internal handpiece components and may lead to blood coagulation inside
the handpiece. Coagulated blood may cause handpiece failure and overheating, causing burn
injuries. To avoid the risk, NSK recommends to immediately lubricate those handpieces with PANA
SPRAY Plus.