Electronic Notifi Cation Setup; Navigating The Website - BeaconMedaes VerusLab 4107 9021 88 Installation, Operation And Maintenance Instructions

Lubricated rotary vane laboratory vacuum system
Table of Contents

Advertisement

1. Click System Trend Setup to access the Sys-
tem Trend Setup page.
2. Select a time interval from the pull-down list
of options.
3. Click the Submit button.
A.7.10 Electronic Notifi cation Setup
This Electronic Notifi cation Setup page (Figure
A.36) is used to confi gure the Electronic Notifi ca-
tion feature of the TotalAlert Embedded control
system. By setting up the Electronic Notifi cation,
key personnel can receive notifi cations of all
alarm/shutdown alerts and/or all maintenance
alerts (e-mail, pager, phone text message). The
device acts as an SMTP client. An SMTP server is
required for electronic notifi cation to function.
Figure A.36 Electronic Notification Setup
1. Click Electronic Notifi cation Setup to access
the Electronic Notifi cation Setup page.
2. Select Enable to enable the e-mail notifi ca-
tion tool.
3. Enter up to fi ve email addresses.
4. For each address, select "System" for that
person to receive all alarm/shutdown alerts.
Select "Maintenance" for that person to re-
ceive all maintenance alerts. An individual
may receive both types of alerts.
5. Enter an email address in the "From" box as
this will be the sender of the notifi cations.
6. If desired, early warning notifi cations can
be set up for maintenance items. By select-
ing these time intervals, a notifi cation will
Lubricated Rotary Vane Laboratory Vacuum Systems
A-17
be sent before the maintenance item is due,
providing time for planning the service activ-
ity.
7. Click the Submit button.
A.7.11 Navigating the Website
The TotalAlert Embedded website allows the user
to easily view the status of all activity pertaining
to the laboratory vacuum system. By clicking the
menu items to the left of the screen, a user can
view pages displaying accurate and timely infor-
mation about the system. These pages include:
Figure A.37 Status Page
Status: The Status page (Figure A.37) displays
current information for both the overall system
(run hours, vacuum level, ambient tempera-
ture, alarms, shutdown, and service) and the
units (run hours, operation mode, run mode, se-
quence, alarms, shutdown, and service).
Figure A.38 Alarms Page
4107 9021 88.01