Lubricated Rotary Vane Laboratory Vacuum Systems
1. Click Device Setup to access the Device Setup
page.
2. Enter the new device name.
3. Enter the location
4. Enter the facility description.
5. Enter the contact information.
6. Click the Submit button.
A.7.7 Network Setup
This Network Setup page (Figure A.33) is used to
confi gure the facility network information and e-
mail server confi guration.
Figure A.33 Network Setup
NOTE:
Obtain the IP Address and DNS Name
from the facility's Information Systems
department.
1. Click Network Setup to access the Network
Setup page.
2. Select Enable or Disable for the DHCP Mode.
3. Enter the IP Address for the DNS server con-
fi guration.
4. Enter the IP Address and DNS name for the
e-mail SMTP server confi guration.
5. Click the Submit button.
4107 9021 88.01
A.7.8 Website Setup
This Website Setup page (Figure A.34) is used to
confi gure the automatic refresh interval for the
website. By selecting an auto-refresh interval,
the web page will be refreshed automatically,
keeping all information current during a viewing
period.
Figure A.34 Website Setup
1. Click Website Setup to access the Website
Setup page.
2. Select a time interval from the pull-down list
of options.
3. Click the Submit button.
A.7.9 System Trend Setup
Figure A.35 System Trend Setup
This System Trend Setup page (Figure A.35) is
used to confi gure the trend log for the website
and to allow the data to be cleared. The System
Trend screen contains a maximum of 720 events
for each of the items recorded, so the actual du-
ration of the trend log changes with the time in-
terval selected. The durations available to select
are 6 hours, 24 hours, 2.5 days, and 10 days.
A-16