Setup Instructions
Perform the following checks before using the bed:
1. Do a careful walk around inspection of the bed. Look underneath the bed to make
sure there is nothing stored there. This area must be kept clear to avoid any
interference with bed parts, which move during operation.
2. When bed is in high position, bed should not be placed against a wall. Completely
raise and lower bed to verify that the bed frame will not contact a wall during use.
3. Lock the casters (two individual lock and two steering control type casters) rotate
steering casters until they lock in place.
4. Plug the power cord into a 100-240 volt AC grounded socket with a capacity of at least
3.9 amperes. Using the handset, cycle the bed through all of its operations. (see page
6) Make sure that everyone (including you) is clear of the bed during these operations.
See that:
a) All powered bed operations work smoothly and stop in the desired position.
b) Bed height, angle of the leg lift and backrest can be adjusted as desired.
5. Position the side rails. Be sure they slide up and down properly.
6. When you are satisfied that the bed is operating properly, instruct all personnel who
may be involved in the bed operation.
If problems are found, contact Gendron customer service at 800-537-2521 or your
authorized Gendron dealer