Safety Checks - Magnet-physik EF 14 Operating Instructions Manual

Electronic fluxmeter
Table of Contents

Advertisement

Electronic Fluxmeter

1.1.5 Safety Checks

All safety checks as described here must be carried out regularly and carefully. Please use the
checklist "General Check" for safety checks
Check Intervals:
 Once a year and
 after every repair
Contents of check:
 specified condition
Table 1: Checklist for safety check
General Check and Functional Tests
General Check:
Switch off the device and disconnect it from mains power supply.
Visual control check of instrument casing
Visual control check of connection cables
Functional test:
Before functional test the device must be connected to mains power supply.
EF 14
Display
EF 14
Display
Check Date:________________
Important!
We recommend having the instrument and, if appropriate, the measuring
coils periodically recalibrated to ensure traceability of the measuring results.
The suitable recalibration interval depends on the operating conditions and of the quality
management stipulations of the owner. Intervals of one to two years are usual, depending
whether the instrument is continuously used in rough production environment or from time to
time under laboratory conditions.
BA – no.: 100514EBA02
 specified function
undamaged
undamaged
switch on with power switch
Display lights up
switch off with "Power Off"-key
extinguishes
Operator (Signature): ________________________
th
December 9
, 2020
EF 14
Page 9