GE
D
GB091046, R
IRECTION
3-6-2-6
Select Units of Measure
Figure 3-12 Select Units of Measure
Table 3-17
STEP
1.
Open the Configuration Window, see
VIVID E9 / VIVID E7 as 'ADM'" on page
Select System, if needed.
2.
Use the drop down dialog to select Metric or US Units.
3.
Restart the scanner.
3 - 28
2
EVISION
Select Units of Measure
TASK
4-2-5 "Logging on to
4-15.
Section 3-6 - Configuration
VIVID E9 / VIVID E7 BT'13 S
UNITS
EXPECTED RESULT(S)
The System Settings window is displayed.
The selected units (Metric or US) will be used for measurements as soon
as the unit has been restarted.
All the changed settings will be used after the restart.
M
ERVICE
ANUAL