Interacoustics VisualEyes 505 Instructions For Use Manual page 33

Hide thumbs Also See for VisualEyes 505:
Table of Contents

Advertisement

Guidance and manufacturer's declaration - electromagnetic emissions
The Instrument is intended for use in the electromagnetic environment specified below. The customer or
the user of the Instrument should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
portable and mobile RF communications equipment and the Instrument.
The Instrument is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Instrument can help prevent electromagnetic interferences
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Instrument as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum
output
power of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
D-0126472-A – 2020/08
VisualEyes™ 505 Instructions for Use -US
Compliance
Group 1
Class B
Not Applicable
Not applicable
Recommended separation distances between
Separation distance according to frequency of transmitter
d = 1.17 √ ����
150 kHz to 80 MHz
0.12
0.37
1.17
3.70
11.70
Electromagnetic environment - guidance
The Instrument uses RF energy only for its internal
function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
The Instrument is suitable for use in all commercial,
industrial, business, and residential environments.
[m]
d = 1.17 √ ����
80 MHz to 800 MHz
0.12
0.37
1.17
3.70
11.70
d = 2.23 √ ����
800 MHz to 2.7 GHz
0.23
0.74
2.33
7.37
23.30