What's In The Kit; Preparation Before Use; Calibration - Apera Instruments PH20 Value User Manual

Ph tester kit
Hide thumbs Also See for PH20 Value:
Table of Contents

Advertisement

2. What's in the Kit?

3. Preparation before Use

3.1. Prepare a cup of pure water (8-16oz) for probe rinsing. Pure water refers to tap water, RO water, distilled
water, or deionized water.
3.1. Pull off the battery insulation paper; Take off the probe cap.
3.2. Perform a 2-point calibration in pH 7.00 and pH 4.00 standard buffers. Refer to Section 4 for tutorial.

4. Calibration

4.1 Short press
to power on.
4.2 Rinse the probe in pure water and shake off excess water.
Long press
to enter calibration mode.
(short press
to go back to measurement mode if you want to cancel
the calibration).
4.3 Insert the probe in the 7.00 pH calibration solution to start 1st point
calibration; Make a quick stir in the solution then hold still;
Wait for
to appear and stay on the screen.
4.4 Short press
to start 1st point calibration; tester returns to
measurement mode;
a successful 1-point calibration.
nd
4.5 To calibrate 2
point, use 4.00 pH buffer and repeat Step 4.2 to 4.4
(Do NOT turn off the tester after you finish pH 7 calibration).
display next to
, indicating a successful 2-point calibration.
4.6 If necessary (target pH>8.0), calibrate 3
pH buffer (sold separately) and repeat Step 4.2 to 4.4.
next to
and
, indicating a successful 3-point calibration.
will display on the lower left corner, indicating
rd
point using 10.01 standard
will
will show up
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PH20 Value and is the answer not in the manual?

Questions and answers

Table of Contents