ETEKCITY EBP-UA5 User Manual page 27

Upper-arm blood pressure monitor
Hide thumbs Also See for EBP-UA5:
Table of Contents

Advertisement

Table 2
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
IEC 60601-1-2 Test level
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
Power supply lines: ±2 kV
input/output lines: ±1 kV
line(s) to line(s): ±1 kV
line(s) to line(s): ±2 kV
100 kHz repetition frequency
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º,
180 º, 225 º, 270 º and 315 º
0% 1 cycle And
70% 25/30 cycles
Single phase: at 0
0% 300 cycle
Compliance Level
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
N/A
N/A
N/A
27