2.18 Setting the High-Level and Low-Level Values

Introduction
To Set the High-
Level and Low-
Level Values
54
You can specify a signal by setting its amplitude and dc offset values.
Another way to set the limits of a signal is to specify its high-level
(maximum) and low-level (minimum) values. This is typically convenient for
digital applications.
In the following example, let us set the high-level to 1.0 V and the low-level
to 0.0 V.
Steps:
Press the Ampl softkey to select Ampl.
Press the softkey again to toggle to show the different
representations of the output voltages.
Press the High Low softkey to select High level and Low level
Both the Ampl and Offset softkeys toggle together, to High and Low,
respectively.
Set the High Level value using the numeric keypad or the knob.
Press the Low Level softkey and set the value using the numeric
keypad or the knob.
These settings (high-level = "1.0 V" and low-level = "0.0 V") are equivalent
to setting an amplitude of "1.0 Vpp" and an offset of "500 mVdc".