Keysight M9380A Troubleshooting

M9380A Troubleshooting

Communications

If you are unable to communicate with the M9380A verify that the following
installations are correct:
If not all modules and their slot locations are visible in the SFP "Connect to
Instrument" dialog:

Module-Level Troubleshooting Overview

Status LED States

38
Keysight IO Libraries Suite
M9380A SFP program
Module and chassis drivers
System Interface Card, cable and PC PXIe card connections, if you are using
an external host PC
Close the SFP.
1.
Start Keysight Connection Expert, by selecting Start > All Programs >
2.
Keysight Connection Expert
still not visible, select
Restart the SFP.
3.
Start by verifying ALL cables are properly connected and SMA connectors are
1.
torqued to 8 Lb-In (0.904 Nm).
page 1
for instrument-level cabling.
Check the front panel Status LEDs.
2.
Check the M9380A SFP Instrument Status window and Self Tests.
3.
Perform a module-level troubleshooting check.
4.
Once you open the M9380A SFP, select all relevant modules, and then
choose Connect. The modules are initialized. Until a module is initialized,
the LEDs are off.
Module
Green
M9300A
Keysight M9380A PXIe Vector Signal Generator 1 MHz to 3 GHz or 6 GHz
. If any or all modules and their slot locations are
Refresh All
.
See "M9380A Instrument Connections" on
See "Status LED States " on page 1
Orange
Red
n/a
Off
Not connected by the SFP.
.