Failure to comply with these instructions could result in death or serious injury and material damage. SIKA, Dr. Siebert & Kühn GmbH & Co. KG, Struthweg 7-9, 34260 Kaufungen, Germany 2/11 Tel: +49-5605-803-0, Fax: +49-5605-803-54, email: info@SIKA.net, web: www.SIKA.net...
If the logger memory contains data already, the menus/parameters marked with (*) can not be invoked! If these should be altered the logger memory has to be cleared before! SIKA, Dr. Siebert & Kühn GmbH & Co. KG, Struthweg 7-9, 34260 Kaufungen, Germany 4/11...
When measuring air temperature the probe has to be dry. Otherwise the cold due to the evaporation causes too low measurings. SIKA, Dr. Siebert & Kühn GmbH & Co. KG, Struthweg 7-9, 34260 Kaufungen, Germany 5/11 Tel: +49-5605-803-0, Fax: +49-5605-803-54, email: info@SIKA.net, web: www.SIKA.net...
Note: The measuring and range values read via interface are always in the selected display unit (°C/°F)! SIKA, Dr. Siebert & Kühn GmbH & Co. KG, Struthweg 7-9, 34260 Kaufungen, Germany 6/11 Tel: +49-5605-803-0, Fax: +49-5605-803-54, email: info@SIKA.net, web: www.SIKA.net...
Please check the settings when necessary. If the battery was replaced the referring menu ‚CLOC‘ will automatically be started. SIKA, Dr. Siebert & Kühn GmbH & Co. KG, Struthweg 7-9, 34260 Kaufungen, Germany 7/11 Tel: +49-5605-803-0, Fax: +49-5605-803-54, email: info@SIKA.net, web: www.SIKA.net...
Calibration certificates can be issued by the factory. For this the device has to be sent to the manufacturer. Just the manufacturer can check the factory settings and correct them if necessary. SIKA, Dr. Siebert & Kühn GmbH & Co. KG, Struthweg 7-9, 34260 Kaufungen, Germany 10/11...