Important! Never autoclave
any component of the
electrophoresis unit or casting
kit.

Care and maintenance

Cleaning
After each use, clean the unit with a mild deter-
gent and water, rinse thoroughly with distilled
water, and allow to air dry. Never use abrasive
cleansers. Do not expose the unit to solutions
or vapors of aromatic or halogenated hydrocar-
bons, ketones, esters, alcohols (over 30%), or
concentrated acids (over 25%).
To reduce DNase and RNase contamination,
soak the buffer chamber or casting kit for 10
minutes in a 3% hydrogen peroxide (H
solution, then rinse thoroughly with DEPC-
treated, autoclaved, deionized water. (Sambrook,
et al. 1:7:40)
Replacing foam pads
Remove worn foam pads. Peel off the adhesive
cover on a new foam pad. Align the pad so
that it will rest on the bottom of the tray along
a short (7-cm) side, adhesive side toward the
inside wall, and then press it in place. Repeat
with second pad on the wall opposite the fi rst
pad.
Replacing the electrode
It is recommended that electrodes be replaced
only by Hoefer technicians. Call your local
representative for advice.
O
)
2
2
p11