Safety Information; Specifications - Fluke 709 Calibration Manual

Precision loop calibrator
Hide thumbs Also See for 709:
Table of Contents

Advertisement

709/709H/710
Calibration Manual

Safety Information

For complete information about how to safely use this Product, read the 709/709H/710
Safety Information included with the Product or located on the Fluke website.

Specifications

Ranges
Resolution
Accuracy ..................................................... 0.01 % ±2 LSD all ranges (@23 °C ±5 °C)
Stability
Operating Temp Range .............................. -10 °C to 50 °C (14 °F to 122 °F)
Storage Temp Range ................................. -20 °C to 60 °C (-4.0 °F to 140 °F)
Altitude ........................................................ 3000 meters
Ingress Protection Rating ........................... IEC 60529: IP40
Humidity Range .......................................... 10 % to 95 % non-condensing
Display
Power .......................................................... Six IEC LR03 batteries
Battery Life (Alkaline Batteries) .................. ≥40 hours continuous use (measure mode)
Loop Compliance Voltage .......................... 24 V dc @ 20 mA
Loop Drive Capability
Electromagnetic Environment..................... IEC 61326-1 (portable equipment)
Dimensions (LxWxD) .................................. 152 x 93 x 44 mm, (6.0 x 3.7 x 1.7 in)
Weight ......................................................... 0.3 kg (9.5 oz)
2
mA ........................................................ 0 mA to 24 mA
Volts ..................................................... 0 V dc to 30 V dc
mA Ranges .......................................... 1 μA
Voltage Range ..................................... 1 mV
20 ppm of F.S. /°C from -10 °C to 18 °C and
28 °C to 55 °C
128 x 64 pixels, LCD Graphic with backlight,
8.6 mm high digits
1200 Ω without HART resistor, 950 Ω with
HART resistor

Advertisement

Table of Contents
loading

This manual is also suitable for:

709h710

Table of Contents