Electronics compartment Ex ia
Service interface (CDI)
Taking the following values into consideration, the device can be connected to the certified
Endress+Hauser FXA291 service tool or a similar interface:
Service interface
U
= 7.3 V
i
effective inner inductance L
= negligible
i
effective inner capacitance C
= negligible
i
U
= 7.3 V
o
I
= 100 mA
o
P
= 160 mW
o
L
(mH) =
5.00
2.00
1.00
o
1)
C
(μF)
=
0.73
1.20
1.60
o
2)
C
(μF)
=
-
0.49
0.90
o
1)
Values according to PTB "ispark" program
2)
Values according to IEC/EN 60079-25, Annex C
0.50
0.20
0.15
0.10
0.05
2.00
2.60
-
3.20
4.00
1.40
-
2.00
-
-
XA00499F-F
0.02
0.01
0.005
0.002
5.50
7.30
10.00
12.70
-
-
-
-
0.001
12.70
-
23