ADC 2200 Instructions For Use page 2

Advantage fingertip pulse oximeter
Hide thumbs Also See for 2200:

Advertisement

Operation Instructions
(Figure 3)
1. Hold the Advantage 2200 with the display screen facing
toward you as shown in Figure 3. Gently squeeze the hinged
end and insert one finger into the device fully. Release hinged
end to allow device to clamp down onto finger. Be sure finger
is fully inserted. (When your finger is inserted into the device,
your nail surface must be facing the same side as the display.)
2. Press the power button on the front panel once.
3. Remain still while taking a measurement and avoid applying
pressure to the clamped device while in use.
4. The device will display a reading once it has had a moment to
begin measurement.
(Figure 3)
Note:
• The rubber inside the oximeter that touches the finger may
be cleaned with alcohol.
• For best results, keep the Advantage 2200 at heart or chest level.
• The index (pointer) finger is recommended.
• While on the finger, do not press the Advantage 2200 against
any surface and do not squeeze it or hold the device.
• The internal spring provides the correct pressure. Additional
pressure may affect readings.
Reading Your Results
After measurement begins, you will see three pieces of information
displayed on your pulse oximeter:
• A pulse bar graph display
corresponding with your
pulse beat will be shown.
The height of the bar
graph shows your
pulse strength.
• Your pulse rate will
be displayed as
BPM or beats per
minute.
• The amount of oxygen in your
blood will be displayed as SpO
Changing Directions of the Display
When you first power on the unit, the display will orient it with the
numbers facing the person wearing the device. The display can be
re-orientated to face the other direction, if desired.
1. To change the direction of the display, press the power button
while the unit is displaying a reading.
Note: By default, the unit will always display the readout facing the wearer.
Maintenance and Storage
1. When the low-power indicator comes on, you must replace
the batteries.
2. The surface of the pulse oximeter that comes into contact with
your finger should be kept clean.
3. When storing the pulse oximeter for more than 30 days, remove
the batteries.
4. When storing the pulse oximeter, ensure that the temperature
is between -13°F to 158°F (-25°C to 70°C) with a relative humidity
less than or equal to 93%.
5. Store the pulse oximeter in a dry environment. Exposure to
moisture may decrease the lifespan of your device.
6. Batteries should be disposed of in accordance with local and
state laws.
Cleaning the Pulse Oximeter
Clean the rubber touching the finger inside of the device with a soft dampened
cloth with 70% isopropyl alcohol, and clean the test finger using alcohol before
and after each test. Do not pour or spray any liquids onto the pulse oximeter,
and do not allow any liquids to enter any openings in the device. Allow the
device to dry thoroughly before reusing.
Press Here
to Open
Pulse
Pulse Bar
Rate
Oxygen
Saturation
Low-Power
Power
Button
Finger Enters
Here
%.
2
Detailed Description of Product Functions
1. Display Type: LED
2. SpO
:
2
Measurement
Range:
70-100%
Accuracy:
70%-100%, ±2%;
0-69% no definition
3. Pulse Rate:
Measurement
Range:
30bpm~250bpm
Accuracy:
30~99BPM, ±2BPM;
100bpm~250bpm, ±2%
Pulse Intensity: Bar Graph Indicator
4. Power Requirements:
Two AAA alkaline batteries
Power
Consumption:
Less than 40mA
Problems
Possible reason(s)
SpO
1. Finger is not
2
or
pulse rate
2. Oxyhemoglobin
cannot be
shown
normally
1. Finger might not
SpO
or
2
pulse rate
is shown
2. Excessive movement
unstably
1. Low battery strength
The pulse
2. Batteries installed
oximeter
cannot
3. The pulse oximeter
be powered
on
1. The product is
Display
suddenly
Graph
powers
off.
2. Low battery strength
Indicator
1. Low power
"Error7"
2. Pulse oximeter is
Limited Warranty
ADC warrants its products against defects in materials and workmanship under normal use and service as follows:
1. Warranty service extends to the original retail purchaser only and commences with the date of delivery.
2. Your pulse oximeter is warranted for one year from date of purchase (all parts).
What is Covered: Replacement of parts, and labor.
What is Not Covered: Batteries where supplied. Transportation charges to and from ADC. Damages
caused by abuse, misuse, accident, or negligence. Incidental, special, or consequential damages. Some
states do not allow the exclusion or limitation of incidental, special, or consequential damages, so this lim-
itation may not apply to you.
To Obtain Warranty Service: Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Dr.,
Hauppauge, NY 11788. Please include your name and address, phone no., proof of purchase, and a brief
note explaining the problem.
Implied Warranty: Any implied warranty shall be limited in duration to the terms of this warranty and in
no case beyond the original selling price (except where prohibited by law).
This warranty gives you specific legal rights and you may have other rights which vary from state to state.
Declaration: This product complies with the IEC60601-1-2 standard. The user-contacting materials are
non-toxic and comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Guidance and manufacturer's declaration - electromagnetic
Guidance and manufacturer's declaration - electromagnetic emission
Emission Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
To register your product visit us at
FOR QUESTIONS, COMMENTS, OR SUGGESTIONS CALL TOLL FREE: 1-800-ADC-2670 OR VISIT
Beijing Choice Electronic Technology Co., Ltd.
2nd Fl., 3rd Fl., and Room 410-412 4th Fl.
No.2 Bldg., No 9 Shuangyuan Rd
Shijingshan District, 100041
Bejing, P.R.C.
Shanghai International
Holding Corp. GmbH (Europe)
EiffestraBe 80, 20537
Hamburg, Germany
IB p/n 93-2200-00 rev 8 (05/05/21)
Low-Power Indication:
Battery Life: Two AAA 1.5V,
Could continuously operate for 18 hrs.
5. Dimensions:
Length:
Width:
Height:
Weight:
(including two AAA batteries)
6. Environment Requirements:
Operation Temperature:
41°F to 104°F
Storage Temperature:
-13°F to 158°F
Ambient Humidity:
Operation:
Storage:
Troubleshooting
1. Reinsert / reposition finger
2. Ensure that sensor is clean
inserted correctly
and not blocked
3. Try taking a reading on a
value is too low to
different finger or warm
be measured
your finger by rubbing
1. Reinsert / reposition finger
be inserted deep
2. Remain still during
enough
measurement
during measurement
1. Replace batteries
2. Reinstall the batteries
incorrectly
3. Contact warranty center
for service
might be damaged
1. Reinsert / reposition finger
automatically powered
2. Replace the batteries
off when no signal
is detected for longer
than 8 seconds
1. Replace batteries
2. Contact warranty center
damaged
for service
emissions for all EQUIPMENT and SYSTEMS
Electromagnetic Environment Guidance
The Pulse Oximeter uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The Pulse Oximeter is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
www.adctoday.com/support/warranty-registration
www.adctoday.com/feedback
0123
1200mAh alkaline.
2.19"
(5.5cm)
1.13"
(2.9cm)
1"
(2.5cm)
1.76oz.
(50g)
(5°C to 40°C)
(-25°C to 70°C)
15% 93%
no condensation
in operation
≤93% no condensation
in storage/transport
Solution(s)
Mfg. for: ADC
®
55 Commerce Drive
Hauppauge, NY 11788
Inspected in the U.S.A.
Made in China
tel: 631-273-9600
1-800-232-2670
fax: 631-273-9659
www.adctoday.com
Printed in China

Advertisement

loading
Need help?

Need help?

Do you have a question about the 2200 and is the answer not in the manual?

Questions and answers