Siemens SITRANS LVL200S Operating Instructions Manual page 26

Dpdt relay
Hide thumbs Also See for SITRANS LVL200S:
Table of Contents

Advertisement

Supplement
- Flanges
- hygienic fittings
Gas-tight leadthrough (optional)
- Leakage rate
- Pressure resistance
- hygienic fittings
Output variable
Output
Turn-on voltage
- Min.
- Max.
Switching current
- Min.
- Max.
Breaking capacity
- Min.
- Max.
Contact material (relay contacts)
Modes (adjustable)
- A
- B
Accuracy (similar to DIN EN 60770-1)
Reference conditions and actuating variables according to DIN EN 61298-1
- Ambient temperature
- Relative humidity
- Air pressure
26
DIN from DN 25, ANSI from 1"
Bolting DN 40 PN 40, Tri-Clamp 1", Tri-Clamp
1½" PN 10, conus DN 25 PN 40, Tuchenha-
gen Varivent DN 50 PN 10
-6
< 10
mbar l/s
PN 64
Bolting DN 40 PN 40, Tri-Clamp 1", Tri-Clamp
1½" PN 10, conus DN 25 PN 40, Tuchenha-
gen Varivent DN 50 PN 10
Relay output (DPDT), 2 floating spdts
10 mV
253 V AC, 253 V DC
10 µA
3 A AC, 1 A DC
50 mW
750 VA AC, 54 W DC
If inductive loads or stronger currents are
switched through, the gold plating on the relay
contact surface will be permanently damaged.
The contact is then no longer suitable for
switching low-level signal circuits.
AgNi or AgSnO and Au plated
Max. detection or overflow/overfill protection
Min. detection or dry run protection
+18 ... +30 °C (+64 ... +86 °F)
45 ... 75 %
860 ... 1060 mbar/86 ... 106 kPa
(12.5 ... 15.4 psig)
SITRANS LVL200S l RELAY (DPDT)