Reaction after fault rec-
tification
SITRANS LVL200S l RELAY (DPDT)
Wrong mode selected
l
à Set the correct mode on the mode switch (max.:
overflow protection, min.: dry run protection). Wiring
should be carried out according to the quiescent
current principle.
? Signal lamp flashes red
Electronics module has detected a failure
l
à Exchange instrument or return instrument for repair
? The signal lamp flashes alternately red and green
instrument defective
l
à Exchange instrument or return instrument for repair
Depending on the failure reason and measures taken, the
steps described in chapter "Set up" must be carried out again,
if necessary.

7.3 Exchange electronics

If the electronics module is defective, it can be replaced by the
user.
In Ex applications only one oscillator with respective Ex
approval may be used.
You find all information to the electronics exchange in the
operating instructions of the new oscillator.
In general, all oscillators of series SW60 can be interchanged.
If you want to use an oscillator with a different signal output,
you carry out the complete setup. You find the necessary,
suitable operating instruction on our homepage.
Note:
Keep in mind that enamelled instrument versions need special
electronics modules. These electronics modules are called
SW60E or SW60E1.

7.4 Instrument repair

If it is necessary to repair the instrument, please contact
Siemens Milltronics Process Instruments Inc. You find the
locations on our homepage "www.siemens.com/processauto-
mation".
Maintenance and fault rectification