9. TECHNICAL SPECIFICATIONS
Automatic turn off:
- After 90 seconds of inactivity before the test (sensor strip inserted into the
meter, drop icon flashing).
- After 60 seconds of inactivity after the test and Error messages Er2,
Er3, Er4, HI and LO.
- After 5 seconds for Error messages Er1, Temperature icon and Battery
icon.
Dimensions:
85.5 mm (L) x 56 mm (W) x 18.2 mm (H)
Weight:
46 g (without batteries)
Meter Operating
- Temperature: 5 - 45 °C (41 - 113 °F).
Conditions:
- Relative Humidity: 20 - 90% (no dew condensation).
Meter Storage
- Temperature: -20 – 50 °C (-4 – 122 °F).
Conditions:
- Relative Humidity: 20 - 90%.
Test strip Storage
- Temperature: 4 - 30 °C (39.2 - 86 °F) (both for unopened vial and after
Conditions:
opening).
- Relative Humidity: 20 - 90% (both for unopened vial and after opening).
Control Solution
4 - 30 °C (39.2 - 86 °F)
Storage
Temperature: