Operating in CI Mode

To Operate the Reagent Gas Flow Control Module

To Operate the Reagent Gas Flow Control
Module
After the system has been switched from EI to CI mode, or vented for any other
C AU T I O N
reason, the MSD must be baked out for at least 2 hours before tuning.
Continuing with CI autotune if the MSD has an air leak or large amounts of
C AU T I O N
water will result in severe ion source contamination. If this happens, you need
to vent the MSD and clean the ion source.
Procedure
1 In the Manual Tune dialog, click the CI tab to access the parameter settings
Figure 49
5977B Series MSD Operating Manual
for controlling the CI gas flow. (See
Manual tune CI tab
Figure
49.)