To View MSD Temperature and Vacuum in Manual Tune

To View MSD Temperature and Vacuum in
Manual Tune
You can also use the GC Control Panel to perform this task. (See
MSD from the 7890 GC Control Panel"
8890 GC Operating Manual.
Procedure
1 In Instrument Control view, select Edit Tune Parameters from the Instrument
2 Click the Values tab to view the MSD Temperatures and Vacuum. (See
Figure 32.
5977B Series MSD Operating Manual
menu to display the Manual Tune dialog.
Figure 32
and
Figure 33
Manual tune values tab
on page 104.) For an 8890 GC, see the
on page 118.)