17
Technical Information

Connection of the High Vacuum Flange

Connection of the High Vacuum Flange
In the case the rotor is suddenly blocked, the torque arising from the
system and the high vacuum flange must be absorbed. Only the
components listed in the following can be used to fasten the pump to
the high vacuum flange.
WARNING!
It is mandatory to connect the pump in such a way to withstand the torque
specified in the "INSTRUCTION FOR USE" section. Specifically pay attention
to clamp design, material of clamps and bolts and bolt fixing torque.
WARNING!
If the pump is fastened on a vacuum chamber with different flange exists the
chance of twisting or tearing-off in case the rotor is suddenly blocked. The
combination of different flange types is not allowed in any case. Agilent
Vacuum will not accept any liability for all damages resulting from not
allowed fastening.
316/332
TwisTorr 804 FS User Manual / 87-901-052-01 (B.00)