Analysis: byte number:0x04(4) 1:03 E9 (1001) 2:00 02 (2)
Pressure value =1001*0.01=10.01
Note: The pressure value read is hexadecimal number, convert it to decimal number,
divided by the decimal point value (0: divided by 1; 1: divided by 10; 2: divided by
100; 3: divided by 1000) and then get real-time pressure values.

9. Responsibility

Within one year from the delivery date, we shall repair or replace the
instrument with any quality fault caused by material parts or our
manufacturing technique free of charge. For non-quality malfunction
during user's operation, we are in charge of repair. But the material cost
and the shuttle transportation fees should be borne by users.