Extend—press and hold this control to extend the width of the bed. Press and hold the control until the control’s indicator is green and you hear a single beep. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Scale Reference—this indicator comes on solid when the bed is correctly in the Scale Reference Position. Zero instructions—shows the sequential steps to zero the scale. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Increase and Decrease controls to adjust the settings. To return all pressure settings to the pressures automatically set by the unit for the set height and weight, press Default. 5. Press Enter to save the pressure settings and return to the Pressure Redistribution screen. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 6
NOTES: Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Clorox Healthcare® is a registered trademark of Clorox Professional Products Company, Inc. Oxycide® is a registered trademark of EcoLab USA, Inc. The UL logo is registered trademark of Underwriter’s Laboratories, Inc. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 8
Virex® is a registered trademark of Diversey, Inc. Wex-Cide™ is a trademark of Wexford Labs, Inc. Auto Contour™, Compella™, FlexAfoot™, Hillrom™, and Quick View™ are trademarks of Hill-Rom Services, Inc. IntelliDrive®, Line-of-Site®, Point-of-Care®, SideCom®, and SlideGuard® are registered trademarks of Hill-Rom Services, Inc.
Contraindication—Use of continuous lateral rotation therapy is contraindicated for patients with cervical or skeletal traction. WARNING: Warning—Do not use the Compella™ Bariatric Bed System with patients who weigh less than 113 kg (250 lb) or more than 454 kg (1000 lb). Patient injury and/or equipment damage could occur. INTRODUCTION This manual contains information necessary for normal operation of the Compella™...
YMBOLS Symbol Description CPR instruction label (on earlier versions of the bed)—shows how to operate the CPR function on the air supply unit (1) and bed frame (2) (page 21) Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 17
Service Required indicator (page 20) Bed battery charge status (page 19) Chair position control (page 29) Bed flat control (page 29) Tilt control (page 29) Reverse tilt control (page 29) Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 18
Bed Up and Down controls (page 28) Bed Not Down indicator—comes on when the upper frame is not in the lowest position (page 28) Head Up and Down controls (page 28) Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 19
Zero control—Scale B (page 47) Magnification Mode control—Scale B (page 48) Frame Setup control and indicator—Scale B (page 46) Raise Bed control and indicator—Scale B (page 46) Scale Reference indicator—Scale B (page 46) Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 20
Bed Exit Alert System—Bed Exiting mode (page 51) Bed Exit Alert System—Out-of-Bed mode (page 48 and page 51) Bed Exit Alert System Volume and Tone Control (page 50 and page Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 21
Head angle (page 21) Patient Controls Nurse Call control (page 53) Head Up and Down controls (page 28) Knee Up and Down controls (page 28) Nurse Call Voice indicator (page 53) Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 22
Shows how to raise the push handles (page 36 or 41) Shows how to stow the push handles (page 36 or 41) Before Transport sequence (page 36) After Transport sequence (page 36) Identifies the manual width adjustment control (page 34) Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 23
Do not store cords here Crush warning: must consult accompanying documents Identifies battery installation location Identifies mains fuse Refer to the bed manufacturer’s user manual for compatible support surfaces (page 54) Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 24
Total bed weight including the safe working load is 950 kg (2094 lb); the bed weight excluding the safe working load is 450 kg (992 lb) minimum Support Surface Shows the support surface dimensions Do not iron Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 25
Symbol Description Hand wash only Bleach with specification Bleach as needed Do not dry clean Do not tumble dry Tumble dry no heat Drip dry Machine wash Safe working load Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 26
EN45501scale) (XX identifies the date of manufacture) (The CE mark was first applied in 2015) Medical Device Black M on green background—signifies the scale (NAWI EN45501 only) is certified to weigh in approved positions) Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 27
(Therapy surface and air supply unit only) According to IEC 60529, Rating for protection against access to hazardous parts by a finger ATTENTION: Consult accompanying documents Manufacturer Date of manufacture Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 28
Do not dispose as Unsorted Municipal Waste WARNING (yellow and black) CAUTION (white and black) Must consult the accompanying documents Alternating current Dangerous voltage Equipment emits electromagnetic energy Protective Earth Equipotentiality Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Use care when you perform medical procedures on the patient. • Warning—Do not use the support surface on a bed frame other than the Compella™ Bed frame. The surface may not fit as intended and could cause a risk of patient entrapment.
Page 30
Warning—Before you plug the power cords in, make sure they are not damaged (cuts, exposed wires, worn insulation, etc.). If either power cord is damaged, do not use the bed. Contact your facility-authorized maintenance person or Hill-Rom Technical Support. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 31
Warning—The communication cable connection on the patient-right side of the air supply unit is to be used to connect to a Compella™ Bariatric Bed frame only. Connecting any other device could cause patient or operator injury and/or cause severe damage to the air supply unit as well as any other incompatible device.
Surface controls Brake and steer controls The Compella™ Bariatric Bed System has these features: • CPR controls at patient right foot • Powered width adjustment (with manual override) • Integrated scale Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Low—the Low (-) indicator flashes when the battery is low. An intermittent tone sounds every 2 minutes when the battery reaches a low charge and the bed is unplugged from AC power. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Correct placement of the patient increases the effectiveness of the SlideGuard® Technology. This is designed to minimize patient migration to the foot end of the bed when you raise the head section. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
When CPR is activated, any controls that are locked out will become unlocked. NOTE: With the foam surface, the use of a CPR board may increase the effectiveness of the CPR. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
The power cord on the patient-right side of the bed supplies power to the bed frame articulation controls North America and charges the built-in battery back-up and IntelliDrive® XL Transport System batteries. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Siderails are intended to be a reminder to the patient of the bed's edges, not a patient-restraining device. When appropriate, Hill-Rom recommends that medical persons determine the correct methods necessary to make sure a patient stays safely in bed. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Because siderails are reinforced for the patient environment, the intermediate siderail may seem heavy. CAUTION: Caution—Do not use the siderails to move the bed. Always push or pull from the headboard or footboard. Otherwise, equipment damage could occur. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
To raise—press and hold the Head Up control until the bed is at the desired height. To lower—press and hold the Head Down control until the bed is at the desired height. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Scale, Three-Mode or Single Mode Bed Exit Alert System, Head-of-Bed Alert and Alert Silence, and alert volume. Instruct visitors not to operate the caregiver controls. They may assist the patient with the patient controls. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
When the bed CPR is activated, any controls that are locked out will become unlocked. Follow your facility’s protocols for lockouts to reduce the likelihood of unauthorized use of the bed controls. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
To lower—press and hold the Knee Down control until the bed is at the desired height. NOTE: To put the bed in an auto contour position, press the Head Up and Knee Up controls at the same time. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
(except for powered width adjustment) to be engaged from any of the caregiver siderail controls except the Lockout control. The battery also powers the nurse call function. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Neutral—to move the bed in any direction, move the pedal to the level position. The neutral position helps with sideway movements in a room or a small enclosed area, or to align the bed with another surface. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Hill-Rom makes no recommendation in regard to the use of physical restraints. Users should refer to legal restrictions and appropriate facility protocols before physical restraints are used. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
When the bed is plugged into AC power, the head angle display is always on. It shows the head angle of the bed unless a weight reading is being taken. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Warning—Make sure there are no objects or devices near or under the bed, especially when the bed width is extended, that may interfere with or contact the bed when lowering. • Warning—Keep the patient within the surface perimeter during width adjustments. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
3. Pull the release lever below the head siderail toward you as you push inward on the head siderail until it stops. 4. Pull the release lever below the intermediate siderail toward you as you push inward on the intermediate siderail until it stops. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Caution—When you lengthen or shorten the foot end of the bed, make sure that the magnets on the surface power cord do not interfere with the movement. • Caution—Do not adjust the bed length when traction equipment is in use. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
The indicator flashes when the battery is low. The indicator will be off if the battery is too low to operate the drive system. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 51
Before transport, make sure that the Bed Up/Down controls are not locked out and operate correctly. You will need to use these controls when the transport is complete. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
10. Adjust the head position to make sure the view is not obstructed from the head end of the bed. 11. Secure all equipment being transported with the bed such as monitors, oxygen tanks, and IV poles. 12. Make sure the push handles are up and locked in position. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Warning—During the transport, use caution so that the bed does not tip or overbalance. 1. Grip one or both of the push handles that are at the head end of the bed. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Turn the unit on. 6. After the transport, adjust the bed width and length as necessary for patient comfort and safety. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Caution—When you lengthen or shorten the foot section of the bed, make sure that the magnets on the surface power cord do not interfere with the movement. Equipment damage could occur. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 56
4. After the transport, adjust the bed width and length as necessary for patient comfort and safety. 5. Optional: To stow the push handles, pull them upward to unlock them, and then swing them inward (toward the center of the bed) into the stowed position. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Warning—Zero the scale before a new patient is put on the bed. Make sure to include standard linens on the bed during the zero process. Failure to do so could cause an inaccurate weight reading; patient injury could occur. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
To weigh—press and release Scale Weigh. On release of the Weigh control, the bed gets the current patient weight. The default configured is to show kg. NOTE: The scale display will flash if the maximum weight is exceeded. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
3. Make sure the bed is not touching anything that could affect the patient weight (other beds or carts, headwall, miscellaneous drainage lines, etc.). Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
If the Brake Not Set alert is sounding, you will not hear the triple beep, so we recommend that you release the brake last. The scale system is now ready for you to zero the scale or weigh the patient. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
To weigh—press and release the Scale control. On release of the Scale control, the display shows the current patient weight in kilograms. At this time, you may set the Brake. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
The Bed Exit Alert System controls are on the flip-up caregiver control pod that is on each intermediate siderail. If your bed is equipped with the Compella™ Low Air Loss Surface with Turn Assist and CLRT (Continuous Lateral Rotation Therapy), Bed Exit should be used in the Normal mode only. It should not be used with the CLRT mode active as it may cause false alerts.
After the patient has returned to the bed, the patient must be returned to the correct position for the Bed Exit mode to re-arm. If the patient is not in the correct position, the system will sound an alert. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
3. Press and hold the Volume control. 4. While you press the Volume control, press the Out-of-Bed control. 5. Press and release the Out-of-Bed control until the desired tone is reached. 6. Clear the alert condition. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
The Bed Exit Alert System controls are on the flip-up control pod that is on the intermediate siderails. If your bed is equipped with the Compella™ Low Air Loss Surface with Turn Assist and CLRT (Continuous Lateral Rotation Therapy), Bed Exit should be used in the Normal mode only. It should not be used with the CLRT mode active as it may cause false alerts.
Press the Enable key until its indicator is on solid, and then press any Bed Exit mode control until the indicator goes off. DJUST THE LERT OLUME 1. The patient must be on the bed. 2. The system must be armed. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
When a Nurse Call control is activated, a signal is sent to the nurses’ station. Voice communication is provided through a speaker/microphone Caregiver Patient that is on the inside of both head-end siderails. Control Control Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Warning—Children, pets, and pests may damage the device and may cause bodily harm to themselves and/or the patient. • Warning—The system is design and intended to be used with a specific Compella™ Bariatric Bed frame. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Hill-Rom for the Compella™ Bariatric Bed System. Use of surface products other than those designed for the Compella™ Bariatric Bed System could substantially reduce the effectiveness of the safety features incorporated into the system.
The bed has two support surface options: • Compella™ Therapy Surface with Turn Assist or Turn Assist with CLRT (see “Compella™ Therapy Surface” on page 59)—the therapy surface system consists of the air supply unit, therapy surface, power cord, and communication cable •...
7. For a foam surface, connect the white, black, and red hoses from the surface to the color-coded connectors on the patient right-hand side of the air supply unit. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 72
If you do not hear a tone, do not use the air supply unit. 13. The Select Attached Mattress Model screen shows. Select the correct surface. With CLRT Without CLRT Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Therapy Surface with Turn Assist • Therapy Surface with Turn Assist and CLRT The Compella™ Therapy Surface supports a patient weight up to 454 kg (1000 lb). The air supply unit has a maximum weight input of 455 kg or 995 lb. NOTE: The weight is entered in 5 kg or 5 lb increments.
Do nothing—Max Inflate mode ends at the end of the 1-minute countdown and the surface returns to the previous mode. If the 10-minute extension is selected, then a 10-minute counter screen shows: Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
2. Press Lck Keypd. The control panel is locked (or unlocked) and the previous therapy screen shows. NOTE: When the control panel is locked, the Lockout indicator shows in the upper right-hand corner of the screen. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
7. When the 30 minutes lapse, the patient returns to the flat position and the display goes back to the Pressure Redistribution screen. Deflate and Inflate the Seat Section for Patient Exit or Entry 1. At the Pressure Redistribution screen, press Mattress Adjust. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Pressure relief is provided when the rotation mode is active. To start CLRT mode, do as follows: 1. At the Pressure Redistribution screen, press Turn & CLRT. 2. Make sure the siderails are raised. Press Setup CLRT. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 78
9. If Max Inflate is selected, the patient is centered and cushions inflated to maximum pressure. From there you can Exit CLRT and return to the Pressure Redistribution screen or you can Resume CLRT. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Deflate mode, Foot Deflate mode, Max Inflate mode, or Transport Deflate mode as necessary. 3. If the side bolsters are deflated for 30 minutes, an alarm sounds, and the Warning: Sides Deflated screen shows: Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Pressure Redistribution screen shows. Deflate the Side Bolsters and Foot Section for Patient Transport To deflate the side bolsters and foot section for patient transport, do as follows: Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
AC power and the air supply unit is correctly connected to the frame (see “Bed Width Adjustment” on page 33). Extend Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
2 minutes, press Inflate and Exit at the Foot Deflating screen. If the Deflate Foot mode has been on for more than 2 minutes, press Inflate Foot Section at Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
This includes single beeps that let the caregiver know these: • Input is necessary to change the functionality of the system. • A full turn has been completed. • Communications with the bed frame has been lost. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Pivoting 3-liter cylinder holder AD298 Adjustable IV pole 4 hooks AD299 One hand adjustable IV pole, 4 hooks AC963 Syringe-driver holder a. Adapter bracket P163 must be installed for use of P158. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Make sure that when you put an oxygen tank in the holder that the tank does not interfere with the head section articulation. To Remove 1. Loosen the thumbscrew that holds the oxygen tank tight in the oxygen tank holder. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
X-Ray machine over the patient. To rotate the arm, pull the T-handle that is on the Patient Helper mount while you turn the arm assembly to the left or right. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
When the pendant is not in use, we recommend that you keep it stored in its slot in the siderail. To Remove Gently pull the pendant out from the siderail. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
1. Loosen the knob on the bottom of the holder bracket. 2. Slide the holder bracket on to the bed mount bracket until it stops. 3. Tighten the knob. 4. Put the holder in the applicable position. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
To install Align the headboard pins with the sockets in the frame and then lower the headboard into the sockets. Push the headboard down until the bottom rests on the frame. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Make sure that the holder is correctly inserted in either equipment socket at the foot end of the bed. • Make sure that the 3-liter cylinder is correctly inserted in the holder. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
As you hold the lower section of the pole, adjust the pole to the applicable height or angle. Make sure that the pole is in a safe position. Tighten the knob. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Clean the fluid spill from the bed system. d. Have maintenance examine the system completely. e. Do not put the bed back into service until it is completely dry, tested, and found to be safe to operate. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Cleaning and disinfection are distinctly different processes. Cleaning is the physical removal of visible and non-visible soil and contaminants. Disinfection is intended to kill microorganisms. Table 1 below summarizes the approved cleaners/disinfectants for use with the associated contact time for disinfection. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Use as many wiping cloths as needed to remove the soil. NOTE: If desired, the Compella™ support surface cover may be removed and laundered to remove visible soil. See “Laundry Guidelines” on page 82. It is important to remove all visible soil from all areas before continuing to remove non-visible soil.
NOTE: If bleach is used with another cleaner/disinfectant, use a new or clean cloth/wipe soaked in tap water to remove any disinfectant residue prior to and after the bleach application. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
7. Install the black filter housing under the side with six connectors. 8. If you have replaced the filters, press Filter has been replaced. This resets the filter counter for another 800 hours. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
(PM) for Joint Commission certification. PM not only meets Joint Commission requirements but can help make sure of a long, operative life for the Compella™ Bariatric Bed System. PM will help minimize downtime due to excessive wear. For a preventive maintenance schedule, refer to the Compella™...
Preventive Maintenance Bed Frame Replace the batteries if any of these conditions occur (refer to the Compella™ Bariatric Bed System Service Manual): • The battery indicator does not come on within 3 minutes of bed connection to AC mains. •...
However, certain components have a short life cycle and will need to be replaced in order for the Compella™ Bed to meet its expected life. These components have a 3-year life expectancy: •...
Air Cells Topper Bladder 3. Make sure the hoses are firmly connected and the air cells are free from leaks. If necessary, contact your facility-authorized maintenance person or Hill-Rom Technical Support. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Different alarms are generated by the air supply unit to notify the caregiver of potentially hazardous conditions. The intended position of the caregiver to respond to alarms or alerts is standing in front of the air supply unit at the foot of the bed. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Do one of these: Timeout about to warning minutes after • Extend Max Inflate. timeout. on the Max Inflate • Stop Max Inflate. display. started. • Let Max Inflate timeout. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
The bed has battery charge or is plugged into AC power. • The bed CPR release handle is in the fully returned position. • The bed width extensions are in the fully extended or fully retracted positions. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
– Topper—on the inside seam, on the patient- right foot corner • Foam surface—on the bottom of the cover, toward the foot end, in the center. Bed Serial Number Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
4" (10 cm) a. Measured from the outermost points (bumpers), excluding air supply unit and mounting bracket. b. Measured from the top of the seat section outer edge to the floor. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 106
50°F to 94°F (10°C to 35°C) ambient only) temperature Relative humidity range 20% to 85% non-condensing Atmospheric pressure 70 kPa to 106 kPa Altitude 3000 m to -330 m (9842.5 ft to -1082.7 ft) Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 107
Sleep deck Support surface 1. There are no user accessible fuses for the bed. Refer to the Compella™ Bariatric Bed System Service Manual (178952) for fuse ratings and replacement procedures. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 108
IEC/EN 60601-1-6 Edition 3.1 IEC/EN 60601-1-8 Edition 2.1 REACH Directive 1907/2006 RoHS Directive 2011/65/EU WEEE Directive 2012/19/EU Degree of protection against electric shock Class I for equipment classification per IEC 60601-1 Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 109
Medical equipment needs special precautions in regard to electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information supplied in the tables that follow. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 110
Harmonic Emissions Class A used for domestic purposes. IEC 61000-3-2 Voltage Fluctuations/ Complies Flicker Emissions IEC 61000-3-3 Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 111
IEC 61000-4-8 characteristic of a typical location in a typical commercial or hospital environment. Note 1: U is the AC mains voltage prior to application of the test level. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 112
RF compliance level above, the P7800 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the P7800. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 113
GSM 1900; 1970 DECT; LTE Band 1, 3, 4, 25; UMTS 2450 2400-2570 Bluetooth Pulse WLAN, 802.11 modulation 217 b/g/n, RFID 5240 5100-5800 WLAN Pulse 802.11 modulation 217 5500 5785 Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
Page 114
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)