The timer's electronic equipment is protected by 4 fuses located on the electronic card.
To replace them, proceed as follows (Fig. 25):
1.
Cut off the power supply
2.
Temporarily remove the timer's guard
3.
Determine the fuse to be replaced (Fig. 26)
4.
Remove the plastic protector (Fig. 27)
5.
Remove the fuse
6.
Replace it with a fuse of same type
Internal protection fuses (5x20) quick
7.
Insert the protection
Fig. 26
8.
Close the timer's guard
9.
Restore the power
®
®
X
X
-
-
M
M
i
i
n
n
d
d
D
D
C
C

15. FUSE REPLACEMENT

POWER VOLTAGE OF THE X-RAY SYSTEM
by removing the adjusting
230 V ± 15 %
8AF – 250V
115 V ± 15 %
12.5AF – 250V