SYSTEM CONTROLLER: AL ARMS
System Controller advisory and hazard alarms are messages that appear on the user interface screen when
some problem with the system needs immediate attention. Some of the alarms indicate problems you can
address yourself (for example, a disconnected power cable that you should reconnect). But most instruct
you to call your hospital contact for instructions.
ADVISORY
SYSTEM
CONTROLLER
ACTIVE
SCREEN
SYMBOLS
14
| System Controller
ALARM MEANS
One of two
power cables is
disconnected
OR
Low battery, power
input is low, with
less than 15 minutes
remaining
System Controller
hardware fault
TO RESOLVE ALARM
1. Promptly connect the disconnected
power cable to the power source
(functioning MPU or two fully charged
HeartMate 3™ LVAD 14 V lithium-ion
batteries).
2. If alarm persists, call your hospital
contact.
1. Promptly connect to a working or
different power source (MPU or two
fully charged HeartMate 3 LVAD 14 V
lithium-ion batteries).
2. If alarm persists, call your hospital
contact.
Call your hospital contact as soon as
possible for diagnosis and instructions.