Please follow these guidelines to ensure mechanical integrity of the fragile shank, electrodes and flex:
Do not touch the shank with your fingers or other objects.
Do not subject the probe to vibrations (potential damage of TiN electrodes and shank). Such
vibrations could arise e.g. when blow-drying the probe with compressed air or N
subjecting the probe to ultrasound.
Always work in a clean environment and wear gloves.
Regularly inspect the probes with an optical microscope to verify that shank and electrodes
are undamaged and clean. Color changes/darkening of the TiN can indicate corrosion or
other damage. Please report such observations.
Do not pinch the flex of the probe and maintain a minimum bending radius of 3 mm.
NEUROPIXELS 1.0 USER MANUAL V1.0.5
Figure 39: Removing the probe from the shipping box.
Figure 40: Removing the probe from the shipping box.