Please follow these guidelines to ensure mechanical integrity of the fragile shank, electrodes and flex:
Do not touch the shank with your fingers or other objects.
•
Do not subject the probe to vibrations (potential damage of TiN electrodes and shank). Such
•
vibrations could arise e.g. when blow-drying the probe with compressed air or N
subjecting the probe to ultrasound.
Always work in a clean environment and wear gloves.
•
Regularly inspect the probes with an optical microscope to verify that shank and electrodes
•
are undamaged and clean. Color changes/darkening of the TiN can indicate corrosion or
other damage. Please report such observations.
Do not pinch the flex of the probe and maintain a minimum bending radius of 3 mm.
•
NEUROPIXELS 1.0 USER MANUAL V1.0.5
Figure 39: Removing the probe from the shipping box.