Display
37
What it means
Error message indicates the battery
voltage lower than limitation.
Error message indicates meter
was not finished calibration.
Error message caused by a
used test strip.
Action
Replace a new battery.
Call customer service for
further information.
Repeat the test with a new
test strip.