Home Aide Easy Trak II User Manual page 21

Table of Contents

Advertisement

The result of the control solution test must be within the range printed on the Test Strip vial label. If
the test result falls outside the specified range printed on the Test Strip vial, repeat the test. You can
perform the Control Solution test using any Calibration Code.
Results falling outside the specified range may be caused by:
→Error in performing the test
→Expired or contaminated Control Solution
→Expired or contaminated
→Improper coding of the
→Malfunction of your
If the problem persists, the system may not be working properly. DO NOT continue to use
the system to test your blood glucose level. If you continue to get results that fall outside the
specified range, please contact your local distributor.
Test Strips
Easy Trak II
Meter
Easy Trak II
Meter
Easy Trak II
Easy Trak II
Blood Glucose Monitoring System
18

Advertisement

Table of Contents
loading

Table of Contents