Decontaminating; High Level Thermal Disinfection; Inspecting - ResMed AirSense S10 Clinical Manual

Hide thumbs Also See for AirSense S10:
Table of Contents

Advertisement

 
5. Release the air outlet by pressing the clip located inside the device.
6. Remove the air outlet by pulling it out through the air outlet socket at the rear of the device.

Decontaminating

Before the disinfection process, each component must be cleaned and rinsed so no visible
contamination is present.
1. Clean all components with a soft bristled brush for one minute while soaking in detergent
solution (see table below). Pay particular attention to all crevices and cavities.
2. Thoroughly rinse each component according to the detergent manufacturer's instructions.
ResMed has tested the following detergents according to the manufacturer's instructions:
Detergent
Alconox (diluted at 1%)
Neodisher MediZym (diluted at
2.0%)

High level thermal disinfection

Part
Cleanable water tub
Air outlet
1. Soak the disassembled components in a hot water bath at 194ºF (90ºC) for 1 minute.
Take care that no air bubbles are trapped against the components.
2. Air dry out of direct sunlight and/or heat.

Inspecting

Perform a visual inspection of all components. If any visible deterioration is apparent (cracking,
crazing, tears, etc), the water tub should be discarded and replaced. Slight discoloration of the
silicone components may occur and is acceptable.
Water temperature
Hot water (approx 140°F or 60°C)
Warm water (approx 113 to 140°F or 45 to 60°C)
Room temperature water (approx 70°F or 21°C)
Warm water (approx 113 to 140°F or 45 to 60°C)
Validated number of cycles
Hot water: 194°F (90°C) for 1 minute.
10
10
Cleanable
Air outlet
water tub



