Setting the safety limit values

If the Time Start/Stop key
is held down while "SAFE" is displayed, the safety
limit values can be changed using the respective up/down keys.
Operator panel and display
Display
Display
Time
Speed
Time
Time
up
Shift key
Start
down
hh:mm
Stop
mm:ss
Tapping the Start/Stop keys or On/Off key starts the particular functions.
The desired default values can be changed using the  or  keys. Use the
shift key
to switch from hour/minute mode to minute/second mode.
If the elapsed time exceeds the value of 100 hours, the display switches from
hour/minute mode
to hour mode
Only whole hours are displayed in hour mode.
The following sections contain detailed information about the individual functions.
Display
Temperature
Speed
Speed
up
Start
Temperature
down
Stop
On
Off

Function Shaking

Press the
key to start or stop the shake function. The speed can be adjusted
during operation. The displayed value flashes until the pre-set speed has been
reached.
When the shaking function is started using button
starts measuring the time until the next switch-off.
Note: The shaking function can only be started when the cover is closed. When
the cover is opened, the shaker automatically switches off the shaking and
heating functions.

Timer

The desired shaking duration is set on the operator panel of the timer using the
Time up/down keys. The shift key can be used to switch from hour mode to
minute mode.
The Time Start/Stop key activates the timer and shaking functions. When the time
is up, the shaking function stops and an acoustic signal is given off. If the cover is
opened before the time is up, the timer goes to "pause" and the display flashes. If the
cover remains open for more than 15 minutes, the timer switches off and an error
Temperature
code appears on the display. The shaker no longer starts when the cover is closed.
up
If the timer is not activated, the Time display automatically shows the operating
down
time following the start of the shaking function.

Function Heating

Once the appliance starts, the display indicates the actual value for the set tem-
perature.
Press the
key to start or stop the heating function. The temperature is entered
using the Temp up/down keys. In operation, the temperature is displayed in 0.1 °C
steps.
The target value appears approx. every 5 seconds and remains on the display for
2 seconds.
, the timer automatically