OHAUS STARTER 3100C Instruction Manual page 15

Bench conductivity meter
Hide thumbs Also See for STARTER 3100C:
Table of Contents

Advertisement

Find Quality Products Online at:
STARTER 3100C Bench Conductivity Meter
The details for the second line are:
Auto EP, 85.1μS/cm, 26.8°C, ATC
|
|
|
|---- Auto Temp. Compensation
|
|
|----------------Temperature value and unit
|
|------------------------- Conductivity value
|---------------------------------------- Auto End Point
The details for the third line are:
Ref.25°C , T.Comp.Coeff. 2.00%/°C
|
|
|
|---------------- Temperature compensation coefficient value
|------------------------------------------- Reference temperature, 20 or 25 °C
The format for the print-out following a TDS measurement is:
End Point, Value, Temp., ATC/MTC
Manual EP, 38.0mg/L, 23.2°C , ATC
Ref.25°C , T.Comp.Coeff. 2.00%/°C
TDS-F 0.5
The format for the print-out following a Salinity measurement is:
End Point, Value, Temp., ATC/MTC
Manual EP, 0.08psu, 25.0°C , MTC
The print-out in case of an error message is:
End Point, Value, Temp., ATC/MTC
Error4
The print-out for a successful calibration is:
Standard: 1413μS/cm
Temperature: 30.0 °C
Calibration value: 1552μS/cm
Cell Constant: 1.66 cm-1
ATC/MTC: ATC
Printing from memory When scrolling through the memory you can print the entry that is currently
viewed by pressing and holding
Recall 08:
EndPoint, Value, Temp., ATC/MTC
Auto EP, 68.8μS/cm, 26.8°C , ATC
Ref.25°C , T.Comp.Coeff. 2.00%/°C