MAINTENANCE

5.1

Error message

Error 0
Memory access error
Error 1
Self-diagnosis failed
Error 2
Measured values out of range
C:
> 199.9 mS/cm
TDS: < 0.1 mg/L or > 199.9 g/L
SAL: > 100.0 psu
Error 3
Measured standard temperature
out of range (5 ... 35 °C)
Error 4
Measuring temperature out of
range (0 ... 100 °C)
Error 9
The current data set has already
been stored once
If there is an error, the meter will beep 3 times to alert.
5.2

Meter maintenance

Never unscrew the two halves of the housing!
The STARTER 3100C does not require any maintenance. To clean, use a damp cloth.
The housing is made of acrylonitrile butadiene styrene (ABS). This material is susceptible to
damage by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any
spillage should be immediately wiped off.
5.3

Self diagnosis

Press and hold
and
Each icon blinks one after the other. This way you may check whether all icons are correctly
shown. The next step is to check that the keys are functioning correctly. This requires user
interaction.
When b b blinks, five icons are displayed. Press the 6 keys in any order. Each time you press a key
an icon disappears from the screen, continue to press the other keys until all the icons have
disappeared.
When the self-diagnosis has been completed successfully, PAS appears. If self-diagnosis fails,
error message Err 1 appears .
GlobalTestSupply
www.
STARTER 3100C Bench Conductivity Meter
Reset to factory settings
Repeat the self-diagnosis procedure and make
sure that you finish pressing all five keys within
two minutes.
Make if the electrode is properly connected
and placed in the sample solution
Keep the standard temperature within the
range for calibration
Check if the electrode is properly connected
and keep the sample temperature within the
range
An endpoint reading can only be stored once.
Perform a new measurement to store.