PASSWORD

Use the PASSWORD (PASSWORD SETUP) menu to set the system password or set the level of
protection that conditionally engages or disengages protection of certain groups of parameters.
USER PASSWORD
Use the USER PASSWORD parameter to enter and change the system and conditional password
used to access protected menus. The PASSWORD is a number between one and four digits. Each
digit of the password can be a number from 0 to 9. The default System Password is 1234. Use the up,
down arrow buttons to enter the new password, and then push ENTER.
PROTECTION
Use the PROTECTION parameter to enable (HIGH) or disable (LOW) password protection for the
conditional parameters. The password is the same as the system password. You can conditionally
password-protect the SOFT CUTOUT and all of the PROGRAM MENU parameters except SELECT
PROG and RUN PROG. Select HIGH or LOW with the left and right arrow buttons and push ENTER to
accept the selection.

CALIBRATE

Calibration parameters must be correct for the Product to function properly.
Use the CALIB (CALIBRATION) menu to access to the calibration parameters for the Product. Access
to the IR target calibration parameters is password protected. Calibration parameters are programmed
at the factory when the Product is calibrated. Qualified personnel can adjust these parameters when
needed to improve the accuracy of the Product.
Do not change the calibration values or control parameters from the factory-set values
unless recalibrating the Product. The correct setting of these parameters is important
to the safe and proper operation of the Product.
The parameters in the CALIB menu are set at the factory and must not be altered unless recalibrating
the Product. Recalibration of the Product should be performed by qualified personnel. The correct
values are important to the accuracy and safe operation of the Product. Access to these parameters is
password protected. In the event that the calibration parameters need to be reentered into the Product,
the constants and their settings are listed in the Report of Calibration shipped with the Product.
CONT SETUP
Use the CONT SETUP (CONTROL SETUP) to access the controller parameters.
TEMP PBAND
The TEMP PBAND parameter is the proportional band in °C that the Product's proportional-integral-
derivative (PID) controller uses for control.
TEMP INT
The TEMP INT parameter is the integration time in seconds that the Product's PID controller uses for
control.