Replace The Battery As Follows - Keithley 2750 Service Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System Service Manual
WARNING

Replace the battery as follows:

1.
Before replacing the battery, refer to the troubleshooting procedures in
Section 4
2.
Remove the Model 2750 top cover using the disassembly procedures in
3.
Locate battery BT100 using the motherboard component layout drawing at the end
of
4.
De-solder the battery from the top side of the circuit board.
5.
Install a new battery, Keithley part number BA-51, taking care to observe proper
polarity.
6.
Solder the battery connection to the circuit board using organic (OA-based) flux
solder.
7.
After soldering, remove all flux using a foam-tipped swab or brush dipped in pure
water. Blow dry the board with dry nitrogen gas, then allow the board to dry for
several hours in a 50°C, low-humidity environment before use.
8.
Reinstall the top cover by following the disassembly procedures in
reverse order.
The precautions below must be followed to avoid personal injury.
Wear safety glasses or goggles when working with lithium
batteries.
Do not short the battery terminals together.
Keep lithium batteries away from all liquids.
Do not attempt to recharge lithium batteries.
Observe proper polarity when installing the battery.
Do not incinerate or otherwise expose the battery to excessive heat
(>60°C).
Bulk quantities of lithium batteries should be disposed of as hazardous
waste.
Dispose of used batteries according to all federal, state, and local laws.