Verifying Frequency; Figure 1-17 Connections For Model 7700 Frequency Verification - Keithley 2750 Service Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

1-34
Performance Verification

Verifying frequency

Follow the steps below to verify the Model 2750 frequency function:
1.
2.
3.
4.
5.
Figure 1-17
Connections for Model 7700 frequency verification
CH1
CH3
CH4
INPUT SENSE CH1
CH2
H L H L H L H L
H L H L H L H L
H L H L
H L H L
H L H L
CH21 CH22
CH11 CH12 CH13 CH14 CH15 CH16
Model 7700
Connect the function generator to the Model 7700 CH1 H and L INPUT terminals
(Figure
1-17).
Install the Model 7700 in Slot 1 of the Model 2750, then turn on the power and
allow the unit to warm up for one hour before proceeding. Be sure the front panel
INPUTS switch is set to the REAR position.
Set the function generator to output a 1kHz, 1V RMS sine wave.
Select the Model 2750 frequency function by pressing the FREQ key. Close Chan-
nel 1 by pressing the CLOSE key, then ENTER for "SINGLE" channel, and keying
in 101.
Verify that the Model 2750 frequency reading is between 0.9999kHz and
1.0001kHz.
CH5
CH6
CH7
CH8 CH9 CH10
H L H L H L H L
INPUT
(V, 2 WIRE)
SENSE
(OHMS, 4 WIRE)
H L H L H L H L
CH17 CH18 CH19 CH20
H L H L
Model 2750 Multimeter/Switch System Service Manual
Function Generator
50Ω
Coax
Cable
Function
Output

Advertisement