Remote Error Reporting; Error Summary; Table B-5 Calibration Errors - Keithley 2750 Service Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System Service Manual

Remote error reporting

Methods to detect and determine the nature of calibration errors are discussed in the fol-
lowing paragraphs.

Error summary

Table B-5
messages.
NOTE
Table B-5
Calibration errors
Error Number
+400
+401
+402
+403
+404
+405
+406
+407
+408
+409
+410
+411
+412
+413
+414
+415
+416
summarizes Model 2750 calibration errors, including error numbers and error
Error numbers and error messages are separated by commas. All calibration
errors will set Bit 3 in the Standard Event Register. See the Model 2750 User's
Manual.
Error Message
"10 vdc zero error"
"100 vdc zero error"
"10 vdc full scale error"
"-10 vdc full scale error"
"100 vdc full scale error"
"-100 vdc full scale error"
"100 2-w zero error"
"10k 2-w zero error"
"100k 2-w zero error"
"10M 2-w zero error"
"10M 2-w full scale error"
"10M 2-w open error"
"100 4-w zero error"
"10k 4-w zero error"
"100k 4-w zero error"
"10M 4-w sense lo zero error"
"100 4-w full scale error"
Calibration Reference