Areas Of Application; Injector Modules; Inserting An Injector Module; Removing An Injector Module - Miele A 300/2 Operating Instructions Manual

Injector modules
Hide thumbs Also See for A 300/2:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 41
Inserting an
injector module
Removing an in-
jector module

Injector modules

 Only place empty, unloaded modules in baskets. Check that
they are correctly engaged before loading them.
Modules must be completely emptied before taking them out.
Placing loaded modules in baskets or removing them can cause
damage to the wash load and cause injury should glassware be
broken.
The injector module is only suitable for use in an A 100 upper basket
or an A 150 lower basket. The injector nozzles must point upwards.
Carefully slide the injector module in and connect it to the water
connection point.
Then press down on the module to secure it on the rails in the bas-
ket.
 Do not hold the basket by the injector module when removing it
from the cleaning machine.
The module could come loose or the water connection could be
damaged.
Release the module from the locking mechanism and pull it out of
the water connection at a flat angle. If the module is removed at too
steep an angle, the lever action may damage the connection point
and the connector.