REPLACEMENT OF FILTER ELEMENT AND SEALS

Warning
For replacement of the filter element or seals, the following procedure is recommended:
Remove protective cover after opening toggle-type fasteners.
Remove filter housing cover as described under 12.2.
Unscrew preliminary filter or sample tube from mounting adapter  (see Figure. 7).
Unscrew mounting adapter.
Unscrew filter knurled screw, inspect filter and replace if necessary.
Inspect filter element seals and replace if necessary.
Inspect O-tings in cover and replace if necessary.
Clean filter space.
Refit filter housing cover in reverse order and insert probe head.
Lock filter housing cover as described under 12.2.
Fit protective cover.
Note
16

PROPER DISPOSAL OF THE DEVICE

At the end of the service life of our products, it is important to take care of the appropriate disposal of
obsolete electrical and non-electrical devices. To help protect our environment, follow the rules and
regulations of your country regarding recycling and waste management.
www.mc-techgroup.com
Prior to carrying out maintenance and repairs, it must be ensured that no
health-endangering contaminants remain in the probe. An appropriate
measure is to flush the probe with inert gas, for example.
Before replacing the filter element, the sample gas supply must be
interrupted!
Graphite seals in combination with the highly heated probe can only be
used a single time. Important to note after fitting the filter housing cover
that the U-bolt is tightened hand-tight in a cold condition and retightened
when the operating temperature is reached.