Check that the filter element is screwed tight and then screw on the mounting adapter again.
•
Screw the preliminary filter or sample tube with appropriate seal into the ¾" thread of the mounting
•
adapter ;
•
Push the filter housing cover with preliminary filter or sample tube into the filter space in the probe
head.
Swing clamp B to the right and with handle C bring eye bolt D in position E, so that the clamp
•
engages in eye bolt D and threaded bolt H; if necessary, push in or pull out the filter housing cover
slightly with the clamping screw A;
Turn handle C into position F and tighten handle A hand-tight by turning clockwise.
•
12.3
CONNECTION OF SAMPLE LINE
Provided on the probe side for connection of the sample line is a ¼" NPT i thread. This thread can be
used for screwing in appropriate connecting unions for lines with the sizes Ø 6 mm (standard), 8 mm or
10 mm.
The sample line is connected as follows:
Loosen the toggle-type fasteners on the insulating cover and hood and remove.
•
Screw the appropriate threaded union into the probe head with sealing tape.
•
Insert the line through the respective opening in the probe base plate and silicone holder.
•
Connect the line to the union. For Swagelok
•
- Fully insert the liner with support sleeve into the union.
- Tighten the union nut finger-tight.
- Prior to tightening, mark the union nut in the 6 o'clock position.
- Grip the body with a spanner and tighten the union nut with 11/4 turns; after a full turn, the mark
must be turned further to the 9 o'clock position.
When connecting hose assemblies to stainless steel unions, a supporting sleeve
must always be used.
The connection must be checked for leaks.
Note
Refit the special hood and close with clamps.
•