12.2

FITTING PRELIMINARY FILTER OR SAMPLE TUBE

The SP2500-H.. probe offers the possibility of fitting or removing the preliminary filter or sample tube in
the process without having to remove the probe head.
For this purpose, the filter housing covers must be removed as follows:
Figure 5
Schematic diagram of filter housing cover
Turn handle A about one full turn anticlockwise, so that the cover is lifted.
Place handle C in position E.
Swing out clamp B to the left (in the direction of G);
Remove the filter housing cover with handle A.
16
SP2500-H | 1.00.02
www.mc-techgroup.com